Trans-cranial Magnetic Resonance-guided Focused UltraSound (tcMRgFUS) system integrated with a 1.5T: an SNR study on DQA phantoms

2018

Background: In recent years transcranial Magnetic Resonance-guided Focused Ultrasound Surgery (tcMRgFUS) treatments have been performed only using 3T units. Since following some internal analysis, planning images obtained using a 1.5T MRI’s body RF coil (as usually done with 3T systems) showed generally reasonable quality in terms of anatomy visualization and SNR but thermal images noise was above acceptable standard for treatments, a dedicated 2-channels head coil was developed (InSightec Ltd.) to ensure an adequate signal-to-noise ratio (SNR). Hence, we present the imaging protocol and the technological methods successfully used with the world-first installation of a certified tcMRgFUS sy…

dose quality assurance phantom focused ultrasound SNR MRISettore MED/37 - NeuroradiologiaSettore MED/36 - Diagnostica Per Immagini E RadioterapiaSettore FIS/07 - Fisica Applicata(Beni Culturali Ambientali Biol.e Medicin)